Documentation

Ozone Monitor Comparison
Which ozone monitor should I purchase? The following table compares the specifications of the various 2B Tech Ozone Monitors and is meant to be helpful for deciding which Ozone Monitor is best for your application. If you have questions or need help with selecting the Ozone Monitor best suited to your needs, please do not hesitate […]

Which ozone monitor should I purchase?

The following table compares the specifications of the various 2B Tech Ozone Monitors and is meant to be helpful for deciding which Ozone Monitor is best for your application. If you have questions or need help with selecting the Ozone Monitor best suited to your needs, please do not hesitate to contact us.

SpecificationModel
202205106-L106-M106-MH106-H211 / 211-GPOM
Measurement PrincipleUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV Absorbance
Dual Beam?Optional UpgradeYesNoNoNoNoYesNo
Federal Equivalent Method (FEM)YesYesYes
N/AN/AN/AYes
Yes
Linear Dynamic Range0-250 ppm0-250 ppm0-100 ppm0-1000 ppm0-10,000 ppm0-20 wt%211: 0-2 ppm
211-G: 0-500 ppb
0-10 ppm
Resolution0.1 ppb0.1 ppb0.1 ppb0.01 ppm0.1 ppm0.01 wt%0.1 ppb0.1 ppb
Precision (1σ for 10-s average; aka rms noise)1.5 ppb  or
2% of Reading
1.0 ppb or
2% of Reading
1.5 ppb or
2% of Reading
0.01 ppm or
2% of Reading
0.05 ppm or
2% of Reading
0.01 wt% or
2% of Reading
0.5 ppb or
1% of Reading
1.5 ppb or
2% of Reading
Accuracy1.5 ppb  or
2% of Reading
1.0 ppb or
2% of Reading
1.5 ppb or
2% of Reading
0.01 ppm or
2% of Reading
0.05 ppm or
2% of Reading
0.01 wt% or
2% of Reading
211: 1.0 ppb or
1% of Reading
211-G: 1.0 ppb or
2% of Reading
1.5 ppb or
2% of Reading
CalibrationNIST TraceableNIST TraceableNIST TraceableNIST TraceableNIST TraceableIodometric TitrationNIST TraceableNIST Traceable
Limit of Detection (10-s measurement, 2σ)3.0 ppb2.0 ppb3.0 ppb0.02 ppm0.10 ppm0.02 wt%1.0 ppb3.0 ppb
Flow Rate (nominal)1 L/min1.8 L/min1 L/min1 L/min1 L/minUp to 20 L/min2 L/min0.8 L/min
Flow Rate Requirement>0.6 L/min>1.2 L/min>0.6 L/min>0.3 L/min>0.2 L/min>0.5 L/min>1.2 L/min>0.5 L/min
Baseline Drift<2 ppb/day
<5 ppb/year
<1 ppb/day
<3 ppb/year
<3 ppb/day
<6 ppb/year
<0.01 ppm/day
<0.03 ppm/year
<0.1ppm/day
<0.3ppm/year
<0.02%/day
<0.05%/month
<1 ppb/day
<3 ppb/year
<2 ppb/day
<5 ppb/year
Sensitivity Drift<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<0.5%/day
<3%/year
<1%/day
<3%/year
Measurement Time, Frequency10 s, 0.1 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz10 s, 0.1 Hz
(fast mode: 2s, 0.5 Hz)
Response Time, 100% of Step Change20 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
20 s
2 points
(fast mode:
4 s, 2 points)
Averaging Times1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
1 min, 5 min,
1 hr
Ozone Unitsppb, pphm, ppm, µg m-3,
mg m-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppm, mg m-3vol %, wt % air,
wt % O2, g m-3,
g Nm-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppb
Pressure Unitsmbar, torrmbar, torrmbar, torrmbar, torrmbar, torrmbar, psi, torrmbar, torrtorr
Temperature Units°C, K°C, K°C, K°C, K°C, K°C,°F, K°C, KK
T and P CorrectedYesYesYesYesYesYesYesYes
Operating Temperature Range0 to 50 °C;
-20 to 50 °C with low T Mods
0 to 50 °C;
-20 to 50 °C with low T Mods
0 to 50 °C0 to 50 °C0 to 50 °C0 to 50 °C10 to 50 °C0 to 50 °C
Operating Altitude~0-13.5 km (0-25 km option)~0-13.5 km (0-25 km option)~0-13.5 km (0-25 km option)~0-5 km~0-5 km~0-5 km~0-13.5 km (0-25 km option)~0-13.5 km (option for higher altitude modifications)
Power Requirement100-240 VAC, 50/60 Hz;
11-14 VDC, nominally 600 mA at 12 V;
7.2 watt
100-240 VAC, 50/60 Hz;
11-14 VDC, nominally 650 mA at 12 V;
8 watt
100-240 VAC, 50/60 Hz;
11-28 VDC, nominally 500 mA at 12 V;
6.0 watt
100-240 VAC, 50/60 Hz;
nominally 500 mA at 12 VDC;
6.0 watt
100-240 VAC, 50/60 Hz;
nominally 500 mA at 12 VDC;
6.0 watt
100-240 VAC, 50/60 Hz;
nominally 250 mA at 12 VDC;
3 watt
211: 100-240 VAC, 50/60 Hz;
11-14 VDC, nominally 1 A at 12 V; 12 watt
211-G: 22 watt
100-240 VAC, 50/60 Hz;
7-12 VDC, nominally 250 mA at 12 V;
3 watt
Size3.7×8.5×11.8 in
9.5×21.5×30 cm
3.7×8.5×11.5 in
9.5×21.5×30 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm
OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm
OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm
OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm
OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm
17×14.5×5.5 in
43x37x14 cm
5.5×3.0x1.5 in
14.0×7.6×3.8 cm
(with GPS)
4x3x1.5 in 10.2×7.6×3.8 cm
(without GPS)
Weight5.5 lb
(2.5 kg)

Without instrument case:
2.8 lb (1.3 kg)
5.8 lb
(2.6 kg)

Without instrument case:
3.1 lb (1.4 kg)
Standard Case:
4.2 lb
(1.9 kg)
OEM Version:
2.8 lb
(1.3 kg)
Standard Case:
3.9 lb
(1.8 kg)
OEM Version:
2.5 lb
(1.1 kg)
Standard Case:
3.9 lb
(1.8 kg)
OEM Version:
2.5 lb
(1.1 kg)
Standard Case:
4.8 lb
(2.2 kg)
OEM Version:
3.3 lb
(1.5 kg)
211: 14.7 lb
(6.7 kg)
211-G: 13.4 lb (5.8 kg)
1.0 lb (0.45 kg)

Without battery:
0.8 lb (0.36 kg)
Data Logger Capacity16,383 lines16,383 lines32,736 lines32,736 lines32,736 lines32,736 lines16,383 lines8,192 lines
Analog Inputs for Internal Logging of Other Instruments,
0-2.5 V
3 Inputs3 InputsNoNoNoNoNoNo
Serial Output, RS232YesYesYesYesYesYesYesYes
USB OutputOptional External AdapterOptional External AdapterYesYesYesYesOptional External AdapterYes
Flash Card OutputYesYesNoNoNoNoYesNo
Data Transfer Baud Rates2400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 1920019200
Analog Voltage Output0-2.5 V0-2.5 V0-2.5 V0-2.5 V0-2.5 V0-2.5 V0-2.5 VOption
Analog Current OutputOptional
4-20 mA
Optional
4-20 mA
4-20 mA4-20 mA4-20 mA4-20 mAOptional
4-20 mA
None
Output RangesUser-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
N/A
DewLine™YesYesYesYesYesNoYesYes
Flow ThroughNoNoOptionOptionOptionYesNoNo
GPS OptionYesYesNoNoNoNoNoIncluded
Bluetooth OptionYesYesYesYesYesYesYesNo
Relay with 2 SetpointsNoNoYes (2)Yes (2)Yes (2)Yes (2)NoNo
Air Pump Lifetime (hours)20,00020,00015,00015,00015,000N/A15,0003,000
Flow MeterYesYesYesYesYesNoYesNo