Documentation

Comparison of 2B Tech Ozone Monitors
Which ozone monitor is right for me? Determine the best ozone monitor for your application using this comprehensive 2B Tech ozone monitor comparison table.

Which ozone monitor should I purchase?

The following table compares the specifications of the various 2B Tech Ozone Monitors and is meant to be helpful for deciding which Ozone Monitor is best for your application. If you have questions or need help with selecting the Ozone Monitor best suited to your needs, please do not hesitate to contact us.

SpecificationModel
202205106-L106-M106-MH106-H211 / 211-GPOM
Measurement PrincipleUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV AbsorbanceUV Absorbance
Dual Beam?NoYesNoNoNoNoYesNo
Federal Equivalent Method (FEM)YesYesYes
N/AN/AN/AYes
Yes
Linear Dynamic Range0-20 ppm0-20 ppm0-20 ppm0-1000 ppm0-10,000 ppm0-20 wt%211: 0-2 ppm
211-G: 0-500 ppb
0-500 ppb
Resolution0.1 ppb0.1 ppb0.1 ppb0.01 ppm0.1 ppm0.01 wt%0.1 ppb0.1 ppb
Precision (1σ for 10-s average; aka rms noise)1.5 ppb  or
2% of Reading
1.0 ppb or
2% of Reading
1.5 ppb or
2% of Reading
0.01 ppm or
2% of Reading
0.05 ppm or
2% of Reading
0.01 wt% or
2% of Reading
0.5 ppb or
1% of Reading
1.5 ppb or
2% of Reading
Accuracy1.5 ppb  or
2% of Reading
1.0 ppb or
2% of Reading
1.5 ppb or
2% of Reading
0.01 ppm or
2% of Reading
0.05 ppm or
2% of Reading
0.01 wt% or
2% of Reading
211: 1.0 ppb or
1% of Reading
211-G: 1.0 ppb or
2% of Reading
1.5 ppb or
2% of Reading
CalibrationNIST TraceableNIST TraceableNIST TraceableNIST TraceableNIST TraceableIodometric TitrationNIST TraceableNIST Traceable
Limit of Detection (10-s measurement, 2σ)3.0 ppb2.0 ppb3.0 ppb0.02 ppm0.10 ppm0.02 wt%1.0 ppb3.0 ppb
Flow Rate (nominal)1 L/min1.8 L/min1 L/min1 L/min1 L/minUp to 20 L/min2 L/min0.8 L/min
Flow Rate Requirement>0.6 L/min>1.2 L/min>0.6 L/min>0.3 L/min>0.2 L/min>0.5 L/min>1.2 L/min>0.5 L/min
Baseline Drift<2 ppb/day
<5 ppb/year
<1 ppb/day
<3 ppb/year
<3 ppb/day
<6 ppb/year
<0.01 ppm/day
<0.03 ppm/year
<0.1ppm/day
<0.3ppm/year
<0.02%/day
<0.05%/month
<1 ppb/day
<3 ppb/year
<2 ppb/day
<5 ppb/year
Sensitivity Drift<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<1%/day
<3%/year
<0.5%/day
<3%/year
<1%/day
<3%/year
Measurement Time, Frequency10 s, 0.1 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz2 s, 0.5 Hz10 s, 0.1 Hz
(fast mode: 2s, 0.5 Hz)
Response Time, 100% of Step Change20 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
4 s
2 points
20 s
2 points
(fast mode:
4 s, 2 points)
Averaging Times1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
10 s,
1 min, 5 min,
1 hr
1 min, 5 min,
1 hr
Ozone Unitsppb, pphm, ppm, µg m-3,
mg m-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppm, mg m-3vol %, wt % air,
wt % O2, g m-3,
g Nm-3
ppb, pphm, ppm, µg m-3,
mg m-3
ppb
Pressure Unitsmbar, torrmbar, torrmbar, torrmbar, torrmbar, torrmbar, psi, torrmbar, torrtorr
Temperature Units°C, K°C, K°C, K°C, K°C, K°C,°F, K°C, KK
T and P CorrectedYesYesYesYesYesYesYesYes
Operating Temperature Range0 to 50 °C;
-20 to 50 °C with low T Mods
0 to 50 °C;
-20 to 50 °C with low T Mods
0 to 50 °C0 to 50 °C0 to 50 °C0 to 50 °C10 to 50 °C0 to 50 °C
Operating Altitude~0-13.5 km (0-25 km option)~0-13.5 km (0-25 km option)~0-13.5 km~0-5 km~0-5 km~0-5 km~0-13.5 km (0-25 km option)~0-13.5 km
Power Requirement100-240 VAC, 50/60 Hz;
11-14 VDC, nominally 600 mA at 12 V;
7.2 watt
100-240 VAC, 50/60 Hz;
11-14 VDC, nominally 650 mA at 12 V;
8 watt
100-240 VAC, 50/60 Hz;
11-28 VDC, nominally 500 mA at 12 V;
6.0 watt
100-240 VAC, 50/60 Hz;
nominally 500 mA at 12 VDC;
6.0 watt
100-240 VAC, 50/60 Hz;
nominally 500 mA at 12 VDC;
6.0 watt
100-240 VAC, 50/60 Hz;
nominally 250 mA at 12 VDC;
3 watt
211: 100-240 VAC, 50/60 Hz;
11-14 VDC, nominally 1 A at 12 V; 12 watt
211-G: 22 watt
100-240 VAC, 50/60 Hz;
7-12 VDC, nominally 250 mA at 12 V;
3 watt
Size3.7×8.5×11.8 in
9.5×21.5×30 cm
3.7×8.5×11.5 in
9.5×21.5×30 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm

OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm

Industrial Version:
16 x 14 x 9.3 in
40.7 x 35.7 x 23.6 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm

OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm

Industrial Version:
16 x 14 x 9.3 in
40.7 x 35.7 x 23.6 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm

OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm

Industrial Version:
16 x 14 x 9.3 in
40.7 x 35.7 x 23.6 cm
Standard Case:
3.6×7.9×9.4 in
9x20x24 cm

OEM Version:
2.5x7x9 in
6.2×17.8×22.9 cm

Industrial Version:
16 x 14 x 9.3 in
40.7 x 35.7 x 23.6 cm
17×14.5×5.5 in
43x37x14 cm


5.5×3.0x1.5 in
14.0×7.6×3.8 cm
Weight5.5 lb
(2.5 kg)
5.8 lb
(2.6 kg)
Standard Case:
4.2 lb
(1.9 kg)

OEM Version:
2.8 lb
(1.3 kg)

Industrial Version:
Approx. 14 lb
(6.3 kg)
Standard Case:
3.9 lb
(1.8 kg)

OEM Version:
2.5 lb
(1.1 kg)

Industrial Version:
Approx. 14 lb
(6.3 kg)
Standard Case:
3.9 lb
(1.8 kg)

OEM Version:
2.5 lb
(1.1 kg)

Industrial Version:
Approx. 14 lb
(6.3 kg)
Standard Case:
4.8 lb
(2.2 kg)

OEM Version:
3.3 lb
(1.5 kg)

Industrial Version:
Approx. 14 lb
(6.3 kg)
211: 14.7 lb
(6.7 kg)
211-G: 13.4 lb (5.8 kg)
1.0 lb (0.45 kg)

Without battery:
0.8 lb (0.36 kg)
Data Logger Capacity16,383 lines16,383 lines32,736 lines32,736 lines32,736 lines32,736 lines16,383 lines8,192 lines
Analog Inputs for Internal Logging of Other Instruments,
0-2.5 V
3 Inputs3 InputsNoNoNoNoNoNo
Serial Output, RS232YesYesYesYesYesYesYesYes
USB OutputOptional External AdapterOptional External AdapterYesYesYesYesOptional External AdapterYes
Flash Card OutputYesYesNoNoNoNoYesNo
Data Transfer Baud Rates2400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 192002400, 4800, 1920019200
Analog Voltage Output0-2.5 V0-2.5 V0-2.5 V0-2.5 V0-2.5 V0-2.5 V0-2.5 VNone
Analog Current OutputNoneNone4-20 mA4-20 mA4-20 mANoneNoneNone
Output RangesUser-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
User-Defined
Scaling Factor in Menu
N/A
DewLine™YesYesYesYesYesNoYesYes
Relay with 2 SetpointsNoNoYes (2x Standard/OEM Versions; 4x Industrial Version)Yes (2x Standard/OEM Versions; 4x Industrial Version)Yes (2x Standard/OEM Versions; 4x Industrial Version)Yes (2x Standard/OEM Versions; 4x Industrial Version)NoNo
Air Pump Lifetime (hours)20,00020,00015,00015,00015,000N/A15,0005,000
Flow MeterYesYesYesYesYesNoYesNo